NAGKT-I electrical instrument test-bed
Also called name
Digital high-precision electrical instrument test bench, high-precision electrical instrument test bench, electrical instrument test bench, digital electrical instrument test bench, digital multimeter calibrator, three-phase instrument test bench calibrator
First. Overview
NAGKT-I digital electrical instrument calibration test bench (also called: digital multimeter calibrator, three-table calibrator) is a five-digit LED AC, DC standard voltage, current generator for Calibration digital meters, a variety of hands-meter and DC 0.2, the exchange of 0.5 below the various types of table head. (0 ~ 12MΩ); AC frequency of 50Hz, 60Hz, 400Hz (digital synthesis, the output voltage of the three-phase voltage range of 0 ~ Button to select). The above items and range conversion are converted by internal switches. Using the latest 16-bit A / D converter chip, conversion speed and accuracy has been greatly improved; AC / DC converter with true RMS value to ensure the accuracy of AC output; linearity is good. Using a new type of composite high-power amplifier, the output power of strong, high reliability, low distortion. Output adjustment using imported potentiometer, long life, high fineness regulation. The machine adopts "closed-loop control", the output stability is high. Voltage short-circuit, open current automatic protection, in full compliance with the new three table identification procedures.
Second, the working principle:
The instrument has a digital synthesis sine wave generator unit, power amplifier unit, power and protection unit, range unit, A / D unit, display unit, keyboard unit and standard resistors and other components.
Third, the technical performance:
2.1 Performance parameters:
	
		
			| 
				Item | 
			
				Test Range | 
			
				Output Range | 
			
				Resolution | 
			
				Max Output | 
			
				  
				Accuracy(23℃±2℃) | 
		
		
			
				DC Voltage 
				  | 
			
				500mV | 
			
				0~550.00mV | 
			
				0.01mV | 
			
				100mA | 
			
				±(Read value×0.05%+max data×0.01%) 
				±(Reading value × 0.05% + full scale value × 0.01%) | 
		
		
			| 
				5V | 
			
				0~5.5000V | 
			
				0.1mV | 
			
				100mA | 
		
		
			| 
				  | 
			
				50V | 
			
				0~55.000V | 
			
				1mV | 
			
				100mA | 
		
		
			| 
				  | 
			
				500V | 
			
				0~550.00V | 
			
				10mV | 
			
				20mA | 
		
		
			| 
				  | 
			
				1KV | 
			
				0~1050.0V | 
			
				100mV | 
			
				10mA | 
		
		
			
				  
				DC Current | 
			
				500uA | 
			
				0~550.00uA | 
			
				0.01uA | 
			
				2V | 
			
				±(Read value×0.1%+max data×0.02%) 
				±(Reading value × 0.1% + full scale value × 0.02%) 
				  | 
		
		
			| 
				5mA | 
			
				0~5.5000mA | 
			
				0.1uA | 
			
				2V | 
		
		
			| 
				50mA | 
			
				0~55.000mA | 
			
				1uA | 
			
				2V | 
		
		
			| 
				500mA | 
			
				0~550.00mA | 
			
				10uA | 
			
				2V | 
		
		
			| 
				5A | 
			
				0~5.5000A | 
			
				100uA | 
			
				1.5V | 
			
				±(Read value×0.1%+max data×0.04%) 
				±(Reading value × 0.1% + full scale value × 0.04%) | 
		
		
			| 
				10A | 
			
				0~11.000A | 
			
				1mA | 
			
				1.5V | 
		
		
			| 
				AC Voltage | 
			
				500mV | 
			
				0~550.00mV | 
			
				0.01mV | 
			
				100mA | 
			
				±(Read value×0.1%+max data×0.05%) 
				±(Reading value × 0.1% + full scale value × 0.05%) 
				  
				AC I 1mA not included 
				  | 
		
		
			| 
				5V | 
			
				0~5.5000V | 
			
				0.1mV | 
			
				100mA | 
		
		
			| 
				50V | 
			
				0~55.000V | 
			
				1mV | 
			
				100mA | 
		
		
			| 
				500V | 
			
				0~550.00V | 
			
				10mV | 
			
				20mA | 
		
		
			| 
				1KV | 
			
				0~1050.0V | 
			
				100mV | 
			
				10mA | 
		
		
			
				  
				AC Current | 
			
				500uA | 
			
				  | 
			
				  | 
			
				  | 
		
		
			| 
				5mA | 
			
				0~5.5000mA | 
			
				0.1uA | 
			
				2V | 
		
		
			| 
				50mA | 
			
				0~55.000mA | 
			
				1uA | 
			
				2V | 
		
		
			| 
				500mA | 
			
				0~550.00mA | 
			
				10uA | 
			
				2V | 
		
		
			| 
				5A | 
			
				0~5.5000A | 
			
				100uA | 
			
				1.5V | 
			
				±(Read value×0.1%+max data×0.01%) 
				±(Reading value × 0.1% + full scale value × 0.01%) | 
		
		
			| 
				10A | 
			
				0~11.000A | 
			
				1mA | 
			
				1.5V | 
		
		
			| 
				Middle Data Resistace | 
			
				Display Data | 
			
				Power | 
			
				Accuracy | 
		
		
			
				10Ω、24Ω、50Ω、100Ω、240Ω 
				500Ω、1kΩ、2.4kΩ、5kΩ 
				10kΩ、24kΩ、10kΩ、24kΩ 
				50kΩ、100kΩ、240kΩ、500kΩ、1MΩ、2.4MΩ 5MΩ、10 MΩ、24 MΩ | 
			
				  
				  
				0.25W 
				  
				  
				  | 
			
				±(Display data×0.1%+10mΩ) 
				±(Reading value×0.1%+10mΩ) 
				  | 
		
		
		
	
2.2: AC distortion: less than 1%.
2.3: DC ripple factor: not more than 0.5%; current 10A not more than 3%.
2.4: Stability: the instrument warm-up after one hour, 1 minute output drift should be less than the full range of the basic range of ± 0.05%, for precision measurements need to warm up 2 hours.
2.5: Overall dimensions: 480 × 340 × 134 (mm).
2.6: Weight: 10Kg.